Specifications :
Chromatography Pump | |
Max pressure | 40MPa |
Flow rate range | 0.001-12.000mL/min |
Flow accuracy | RSD≤0.05% |
Flow repeatability | RSD≤ 0.05% |
Pressure display accuracy | ≤0.001Mpa |
Column Oven | |
Temperature range | Room temperature +5°C to 85°C |
Temperature control accuracy | ±0.01℃ |
Temperature stability | ±0.1°C |
Conductivity detectors | |
Pool volume | ≤0.6µL |
Resolution | ≤0.00238ns/cm |
Detection Range | 0~15000μs/cm |
Range | 4 Stage |
Baseline noise | ≤0.05%FS |
Baseline drift | ≤0.4%FS |
Linear range | ≥102 |
Temperature control range | Room temperature +5°C to 75°C |
Temperature control accuracy | ±0.01℃ |
Autosampler | |
Repeatability | Full-loop injection RSD < 0.3 %, Partial injection RSD < 0.5 %, Nondestructive injection RSD < 1 % (injection volume > 5uL). |
Linearity | >0.999 |
Cross-contamination | <0.01% |
Number of sample positions | 120 (2*60,2mL) |
syringe size | Standard: 500uL; optional: 250uL, 1000uL, 2500uL. |
Dosing ring specifications | Standard: 100uL; optional: 20uL, 50uL, 200uL. |
Inlet valve switching time | <60ms |