Fison Instruments Ltd

Atomic Absorption Spectrophotometer FM-AAS-A101

Features:

  •   Integrated flame/graphite furnace atomization system, changeable with flame emission burner
  •   Turret for 8 pre-aligned Hollow Cathode Lamps fitted with pre-heating, and individual pre-centered lamps
  •   Spectral bandwidth of 0.1, 0.2, 0.4, 1.0 and 2.0 nm
  •   Characteristic concentration of copper: 0.02 μg / ml / 1 %
  •   Detection limit of copper: 0.004 μg / ml
  •   PID and Dual curve mode light-controlled temperature control technique ensures fast heating and analytical sensitivity
  •   Fully automated wavelength scanning and peak searching
  •   Advance software for complete parameter settings
  •   Reliable fully automatic graphite furnace analysis
  •   Alarm feature to protect fuel gas leakage, abnormal flow, insufficient air pressure and abnormal flame extinction in flame system
  •   Analysis of elements by flame absorption method and emission method
  •   Strong database system which possesses more than 500 data self-storage
  •   Results printed as parameter, data result and diagram

Specifications:

Wavelength range  190 to 900 nm
Wavelength accuracy ± 0.15 nm
Wavelength repeatability ≤ 0.04 nm
Spectrum bandwidth 0.1, 0.2, 0.4, 1.0 and 2.0 nm
Base line stability ≤ 0.002 Abs / 30 min
Grating 1800 lines / mm
Inflamer All-metal titanium burner
Atomizer Efficient glass atomizer
Lamp Hollow cathode lamp
Lamp stand 8 lamps automatic turret, automatic alignment
Characteristic concentration of copper 0.02 μg / ml / 1 %
Detection limit of copper 0.004 μg / ml
D2 Background calibration ability 1 A ≤ 30 times correction ability
Measurement Repeatability 0.005
Flame system Air-acetylene flame system
Acetylene air burner 100 mm
Ignition dynamic baseline drift ≤ 0.006 Abs / 30 min
Characteristic viscosity ≤ 0.025 μg / ml / 1 %
Standard deviation of the accuracy ≤ 0.5 % (Cu, absorbance > 0.8 A) (detection limit Cu ≤ 0.008 μg / ml)
Protection system Can inevitably cut off gas if pressure is low, power cut-off, flame out and alteration of the burner
Power supply 220 V ± 22 V AC
Dimension 700 × 550 × 450 mm
Net weight 75 kg

Catalogue: PDF

Brand

Fison Instruments Ltd

Fison Instruments Ltd