Specifications :
| Ion Chromatographic Pump: | |
| Type | Self-Regenerating electrolytic micro-membrane suppressor |
| Maximum Pressure | 6.0 Mpa |
| Pressure Display Accuracy | ≤ 0.1 MPa |
| Flow Range | 0.001 ~ 9.999 mL/min |
| Pressure Fluctuation | ≤ 0.5% |
| Flow Stability | (0.2-0.5) mL/min ≤ 2%; (0.5-1.0) mL/min ≤ 1%; > 1.0 mL/min ≤ 1% |
| Manual Sample Injector: | |
| Contact Material of the Rotor | PEEK |
| Contact Material of Medium | PEEK/Ceramics |
| Column Heater: | |
| Operating Temperature Range | Room temperature +5°C~60°C(41~140°F) |
| Controlling Temperature Accuracy | ±0.01°C |
| Allowable Deviation of Column Heater’s Temperature | ± 1°C |
| Temperature Stability | ≤ 0.05°C/h |
| Conduction Detection System: | |
| Cell Volume | ≤0.8μL |
| Detection Range | 0~35000 μS/cm |
| Detection Resolution | ≤0.0020nS/cm |
| Output Voltage | -6000~+6000 mv (adjustable) |
| Electronic Noise | 0.02 nS |
| Baseline Noise | ≤ 0.001 μS/cm |
| Baseline Drift | ≤ 0.02μS |
| Temperature Compensation | 1.7 %/°C |
| Linear Range | ≥ 10 3 |
| Instrument Linearity | ≥0.999 |
| Quantitative Repeatability | ≤1.0% |
| Qualitative Repeatability | ≤0.5% |
| Minimum Detectable Concentration | Cl- ≤ 0.0005 ug/mL; Li+ ≤ 0.001 ug/mL; BrO3≤ 0.001 ug/mL |
| Flow System: | |
| Plastic Flow Path | Made of PEEK materials |
| Six-way Valve | PEEK material, pressure 5000 psi; Independent automatic collecting and flow function. |
| Suppressor: | |
| Dead Volume | <50 μL |
| Other Specifications: | |
| Dimension (LxWxH) | 350x470x510 mm |
| Net Weight | 26 kg |
| Gross Weight | 32 kg |
| Power | 150 W |
Catalogue : PDF






Reviews
There are no reviews yet.