OPTIZEN™ NanoQ

Specification:

Product Qualification

Product Name OPTIZEN NanoQ Plus OPTIZEN NanoQ
Photometrics System Microvolume Spectrophotometer
Light Source(s) Xenon flash lamp
Lifetime Up to 10 years
Detector CMOS linear image sensor (2048 pixels)
Spectral Bandwidth 1.0 nm (FWHM at Hg 253.7 nm)
Wavelength Range 190 – 850 nm
Wavelength Display (setting) 1 nm
Wavelength Accuracy ± 1 nm
Microvolume
Minimum Sample Volume 1 μL
Photometric Range 0.02 – 330 A (10mm equivalent)
Detection Limit (Microvolume) 2 ng/μL (dsDNA)
0.06 mg/mL (BSA)
0.003 mg/mL (IgG)
Maximum Concentration 16,500 ng/μL (dsDNA), 400 mg/mL (BSA)
Cuvette
Photometric Range 0 – 2 A
Detection Limit 0.2 ng/μL (dsDNA)
0.006 mg/mL (BSA)
0.0003 mg/mL (IgG)
Center Height (Z-height) 15 mm
Heating (Optional) 37 °C
Absorbance Precision 0.002 A (0.5 mm path) or 1%
Absorbance Accuracy 3% (at 0.97A at 302 nm)
Measurement Time < 8 seconds
Software Compatibility Windows® 7 and 10
Touchscreen Multipoint capacitive touch
CPU Octa Core ARM® CortexTM-A53 Processor
Storage 32 GB Internal Storage
Glove Compatibility Compatible with lab gloves
Connectivity 4 x USB ports, Ethernet, and RS-232
Display 7-inch, 1280 x 800 HD color display
Operating System (OS) AndroidTM
Footprint (W*D) 216*290 mm
Weight 3.0 kg

 

  • Product details

    There is no need for a user manual since it provides a simple and intuitive user environment. The designated factor is set as the default value for each mode, and the input mode is provided to allow the users to set a factor value which is different from the default. Users can specify the factor ppropriate for the user preference, the lab environment, and the type of protein.

    Measurement menu configuration
    Menu Factor
    Nucleic Acid
    (ng-㎝/㎕)
    dsDNA 50
    ssDNA 33
    RNA 40
    miRNA 33
    Custom Input
    Protein
    (g-㎝/㎕)
    BSA 1.5
    SA 1.49, 1.72
    IgG 0.71, 0.74
    IgE Human 0.65
    Lysozyme 0.38
    OD1 1
    OD600 OD600 1
  • Full Spectrum Analysis
    • OPTIZEN NanoQ can measure the absorption spectrum UV-VIS (190 nm~850 nm) in seconds using by array type spectrophotometer technology with Xenon lamp and CMOS. sensor(2048 pixels). It also provide various algorithm like Peak/Valley detection.

    Catalogue: PDF

Reviews

There are no reviews yet.

Be the first to review “OPTIZEN™ NanoQ”

Your email address will not be published. Required fields are marked *

Scroll to Top