Specifications :
| Product Name | Ambient Air Quality Monitoring Instruments AQ7030 |
| Test Item | CO |
| Test principles | Non-dispersive infrared & gas correlation filtering |
| Range | 0~50 ppm |
| Zero Noise | ≤0.25 ppm |
| Minimum Detection Limit | ≤0.5 ppm(2σ) |
| Range Noise | ≤1 ppm |
| Error of Measurement | ±2% F.S. |
| 20% Range Precision | ≤0.5 ppm |
| 80% Range Precision | ≤0.5 ppm |
| 24h Zero Drift | ±1 ppm |
| Voltage stability | ±1% F.S. |
| Flow stability | ±10% |
| Linear | ±1% F.S. |
Catalog : PDF