Specifications :
Physical Specifications |
|
Power Requirements | 220V,50HZ |
Ambient Temperature | 4-40℃ |
Ambient Humidity | 20-85% relative humidity(non-condensing) |
System features |
|
Qualitative Repeatability(Retention time) | RSD≤0.3% |
Quantitative Repeatability(Area) | RSD≤0.6% |
Pump |
|
Flow Rate Range
0.0001-10.000mL/min( increments :0.0001 mL/min) |
|
Max Operating Pressure | 42MPa |
Pressure Pulsation | <0.1MPa( Test condition: 10MPa , at 1.0ml/min methanol) |
Flow Accuracy | ±0.2%( Test condition: 10MPa , at 1.0ml/min methanol) |
Flow Precision | RSD≤0.06%( Test condition: 10MPa , at 1.0ml/min methanol) |
Gradient Accuracy | ±1% |
Column Oven |
|
Temperature Setting Range | 5 °C ~ 85 °C. |
Temperature Stability | ≤0.2°C/h |
Temperature Controlling Precision | ±0.1 °C |
Autosampler |
|
Injection Volume Precision | <0.25%RSD( full loop) <0.5%RSD( partial loop) <1%RSD( no waste) |
Linearity | >0.999 |
Cross Contamination(Carry Over) | <0.01% |
Sample Capacity | 120 Vials |
Syringe | 250 μL( Standard, with 500μL,1000μL, 2500μL optionals |
Sample Loop | 20μL( Standard, with 50μL, 100μL, 200μL optional) |
DAD Detector |
|
Noise | 1.5×10-5Au |
Drift | ≤1×10-3Au/h |
LOD | 5.0×10-9g/mL (naphthalene/methanol) |
Wavelength Range | 190-800nm |
Wavelength Accuracy | ≤±1 nm |
Wavelength Reproducibility | ≤±0.1 nm |
Linearity | ≥1.0×104 |
Catalog : PDF