Specifications :
PARAMETERS
21 parameters : WBC, Lymp%, Mid#, Gran#, Lymp%, Mid%, Gran%, RBC, HGB, MCV, MCH, MCHC, RDW-CV, RDW-SD, HCT, PLT, MPV, PDW, PCT, P-LCR, P-LCC. 3 histograms for WBC, RBC and PLT.
PRINCIPLES
Impedance method for cells counting. Cyanide free colorimetric method for hemoglobin measurement.
THROUGHPUT
60 samples per hour.
MEASURING CHANNELS
Double channel for counting.
REAGENTS
Diluent 360.20, Lyse 360.5, Cleaner 360.05
SAMPLE VOLUME
Whole blood : 14 µL,
Pre-diluted blood : 20 µL.
REFERENCE VALUE SETTINGS
Preset five sets of reference values, man, woman, child, baby and general.
REPEATABILITY
WBC ≤ 2.0 %
HGB ≤ 1.5 %
MCV ≤ 0.5 %
RBC ≤ 1.5 %
PLT ≤ 4.0 %
PRINTER
Built-in patented structure thermal printer, paper 80 mm wide, easy to load paper. Report printing takes only 5 seconds.
MEASURING OF UNCLOGGING
Black flushing, soaking and burning. Automatic unclogging treatment before and after the sample analysis.
LANGUAGES
English, French, other languages on request.
STORAGE
Can store more than 10 000 sample results, including parameters, histograms, upgradable on request.
COMMUNICATION
Open communication protocol, support connection to LIS.
DATA PORTS
One network port, one RS-232 port, 4 USB ports, Import or export data via USB port, Upgrading software via USB port, Keyboard, mouse, barcode scanner, and multiple printers connected via USB port.
QUALITY CONTROL SYSTEM
Inlude 9 QC files, automatically draw quality control chart, provide the average, SD, CV.
CALIBRATION SYSTEM
Support manual and/or automatic calibration.
DIMENSIONS & WEIGHT
368mm x 448mm x 475mm and 23 kg.
OPERATING ENVIRONMENT
10°C – 30°C temperature and 70% humidity
Catalogue : PDF
Reviews
There are no reviews yet.