Temperature |
10~30 ˚C |
Humidity |
30%~ 80% |
Sample System |
Samples |
192 samples (Extended up to 384) |
Pipette capacity |
10 to 1000µl |
Dispensing |
one aspirating and dispensing probe |
|
Reagent System |
Reagent racks |
Universal- standard 6 positions (extended up to 32); Reagent bottle: Standard 20 reagent bottles position |
Pipette capacity |
10 to 1000 µl |
Pipette precision |
(100 µl) CV≤0.5% |
Time of Dispensing |
5mins (96 well microplates) |
|
Incubation system |
Incubation |
Two independent incubators |
Temperature range |
Room Temperature 45 ˚C |
Temperature Accuracy |
± 0.5 ˚C |
|
Wash Unit |
Probes |
2line- eight nozzle manifolds |
Containers |
Two wash buffers 2L, 1 Distilled water, 1 Waste water |
Residual |
≤2 µl |
|
Reading |
Channels |
8 independent photometric channels |
Reading |
Monochromatic or dichromatic reading |
Absorbance |
0.0 to 3.0 OD |
Spectral range |
400-700nm |
Optical Filters |
2 Standard filters (450nm, 630nm) |
|
Other Specifications |
Software |
Windows 7 |
LIS system |
Bi-direction, support HL7 protocol |
Results |
Absorbance and results are viewed by software |
Printer |
External Optional |
Power Consumption |
400W |
Power Supply |
AC220V ±10%, 50 Hz; 110V ±10%, 60H |
External Dimension |
933 x 685 x 860 mm |
Net Weight |
130 kg |
Gross weight |
180 kg |
|
Reviews