Specifications :

Water Inlet Tap water: TDS<200 ppm (Extra pretreatment filter is recommended, if TDS>200 ppm)
Temperature 5-45°C
Pressure 1.0-4.0 Kgf/cm²
Flow Procedure** PF+AC+RO+AC+DI+TF
Ion rejection rate 96%-99% (New RO membrane)
Organic rejection rate >99%, when MW>200 Dalton
Particles and bacteria rejection rate >99%
Bacteria <0.1 cfu/ml
Output(25°C)**** 45 L/hr
Pure water outlet RO water and Deionized water
DimensionLxWxH 640×540×1110 mm
Weight 70 kg
Standard configuration Main body (Including 1 set of cartridges)+ 2 built-in15 liters tank +accessory bag
Power Consumption (W) 120 W
Power Supply AC110-220 V, 50/60 Hz
Note *The feed water quality will influence the pure water´s quality and cartridges´ life-span. **PF:polypropylene spun fiber, AC:active carbon, RO:reverse osmosis, DI:ion exchange, TF:terminal microfiltration. ***All the specifications are tested under the situation:feed water´s TDS=200ppm, 25°C, 50psi and 15% recovery rate.
Resistivity >10-18.2 MΩ.cm
Conductivity 0.055-0.1µs/cm
Particle(>0.2µm) <1/ml