Medium Water Purification System BDPS-207

Specifications :

Feed Water Requirements*:
      Water Inlet Tap water: TDS<200 ppm (Extra pretreatment filter is recommended, if TDS>200 ppm)
      Temperature 5-45°C
      Pressure 1.0-4.0 Kgf/cm²
Flow Procedure** PF+KDF+AC+RO+UV+DI+TF
Ion rejection rate 96%-99% (New RO membrane)
Organic rejection rate >99%, when MW>200 Dalton
Particles and bacteria rejection rate >99%
Bacteria <0.1 cfu/ml
Output(25°C)**** 30 L/hr
Pure water outlet RO water and Ultrapure water
Water Quality Monitor
DimensionLxWxH 500×360×540 mm
Weight 25 kg
Standard configuration Main body (Including 1 set of cartridges)+15 liters tank+ accessory bag
Power Consumption (W) 120 W
Power Supply AC110-220 V, 50/60 Hz
Note *The feed water quality will influence the pure water´s quality and cartridges´ life-span. **PF:polypropylene spun fiber, KDF:kinetic degradation fluxion, AC:active carbon, RO:reverse osmosis, DI:ion exchange, UV:ultraviolet, UF:ultrafiltration, TF:terminal microfiltration. ***Value of number will be influenced by temperature and feed water quality. ****All the specifications are tested under the situation:feed water´s TDS=200ppm, 25°C, 50psi and 15% recovery rate.
Ultrapure Water Quality:
      Resistivity(25°C) 18.2 MΩ.cm
      Heavy Metal Ion <0.1 ppb
      TOC*** <3 ppb
      Particle (>0.2µm) <1/ml
      Flow rate 2.0 L/min (with pressure tank)

Catalogue : PDF