Open Path H2O Analyzer

Specifications:

CO2 Measurements

    • Calibration range: 0 to 3000 µmol mol-1
    • Accuracy: Within 1% of reading
    • Zero drift(per °C):
      • ±0.1 ppm typical
      • ±0.3 ppm maximum
    • RMS noise (typical @ 370 ppm CO2):
      • @5 Hz: 0.08 ppm
      • @10 Hz: 0.11 ppm
      • @20 Hz: 0.16 ppm
    • Gain drift (% of reading per °C @ 370 ppm):
      • ±0.02% typical
      • ±0.1% maximum
    • Direct sensitivity to H2O (mol CO2 mol-1 H2O):
      • ±2.00E-05 typical
      • ±4.00E-05 maximum

H2O Measurements

    • Calibration range: 0 to 60 mmol mol-1
    • Accuracy: Within 1% of reading
    • Zero drift (per °C):
      • ±0.03 mmol mol-1 typical
      • ±0.05 mmol mol-1 maximum
    • RMS noise (typical @ 10 mmol mol-1 H2O):
      • @5 Hz: 0.0034 mmol mol-1
      • @10 Hz: 0.0047 mmol mol-1
      • @20 Hz: 0.0067 mmol mol-1
    • Gain drift (% of reading per °C @ 20 mmol mol-1):
      • ±0.15% typical
      • ±0.30% maximum
    • Direct sensitivity to CO2 (mol H2O/mol CO2):
      • ±0.02 typical
      • ±0.05 maximum

General

  • Fundamental Gas Sampling Rate: 150 Hz
    • Bandwidth: 5, 10, or 20 Hz; software selectable
    • Type: Absolute, open-path, non-dispersive spectroscopy
  • Detector: Thermoelectrically cooled lead selenide
    • Path Length: 12.5 cm (4.92″)
  • Air Temperature Thermistor: 10K ohm @ 25 °C thermistor
    • Measurement Range: -40 to 70 °C
    • Sensor Accuracy: ±0.25 °C from -20 to 70 °C
    • Resolution: 0.003 °C @ 25 °C
  • Pressure Sensor:
    • Measurement Range: 20 to 110 kPa
    • Sensor Accuracy: ±0.4 kPa from 50 to 110 kPa
    • Resolution: 0.006 kPa
  • Outputs: Ethernet
  • Operating Temperature Range: -25 to 50 °C (-40 to 50 °C verification on request)
  • Relative Humidity Range: 0-95% (non-condensing)
  • Weatherproof Rating: Tested to IEC IP65 standard user
  • Interface: Windows® PC software
  • Power Requirements: 10.5 to 30 VDC
  • Steady-State Power Consumption: 4 W typical at 25 °C
    8 W max over operating range of -25 to 50 °C
  • Head:
    • Size: Diameter 6.5 cm, Length 30 cm
    • Weight: 0.67 kg; 1.3 kg with mounting post
    • Head Cable Length: 200 cm
  • DSI Box:
    • Size: 13.24 × 14.64 × 6.24 cm (H × W × D)
    • Weight: 0.93 kg

Catalogue: PDF

Reviews

There are no reviews yet.

Be the first to review “Open Path H2O Analyzer”

Your email address will not be published. Required fields are marked *