Haze Meter YH1200

Features :

1.Double standard ISO & ASTM

Able to meet the test standard requirements of different users, in accordance with ASTM D1003/1044,ISO 13468,ISO 14782,GB/T 2410,JJF 1303-2011,CIE 15.2,JIS K7105,JIS K7361,JIS K 7136.

2.Easy to operate, faster and more accurate measurement

The YH1200 haze meter is equipped with a large-size touch screen for easy operation.With a PD array detector, CIE V(λ)2 degree visual response enables high precision and repeatable transmittance and fog measurements.USB data output device for docking with laboratory system.

3.Dynamic measurement

Independent light source detector and temperature sensor, constantly monitor light source and environmental change, ensure the reliability of test data.

4.With compensation port, so that the measurement data more accurate

Meet ASTM D1003 non-compensation method, ISO 13468 compensation method, full light transmittance, fog test, provide more accurate test results.

5.Easy to measure and widely applicable to samples

Open measuring area, vertical and horizontal testing, suitable for more samples to be tested.

6.Quality control software

It provides powerful software for measuring and analyzing fog and light transmittance, which is suitable for quality monitoring and tabulated management of fog and light transmittance data in various industries.The management of users will be digitized at the PC end, the difference of fog and light transmittance will be compared, and the test report form will be generated to facilitate customer customization and management.

SKU: 3nh-YH1200-Haze Meter Category: Tags: , , , , , ,

Specifications :

Product Model  YH1200 Haze Meter
Optical Geometry Transmittance:0/D(Parallel light illumination, diffuse viewing)
Conforms to:ASTM D1003/1044,ISO 14782,GB/T 2410,JJF 1303-2011,CIE 15.2,JIS K7105,JIS K7361,JIS K 7136
Character The instrument can easily achieve ASTM D1003 non-compensation method, full light transmittance,haze test. Open measuring area for vertical and horizontal testing. In glass processing, plastic processing, film, screen processing, packaging and other industries are widely used in transmittance,haze detection.
Integrating Sphere Size Φ154mm
Light Source 400nm to 700nm, combined LED light
Specular Component /
Sensor The PD array detector satisfies CIE V(λ)2 degree visual response
Wavelength Range /
Wavelength Interval /
Semiband Width /
Transmittance measurement range 0~100%
Measuring Aperture Φ20mm/Φ15mm/Φ8mm/Φ4mm(Select a single aperture)
Sample Size Thickness <170mm
Color Space /
Color Difference Formula /
Other Colorimetric Index Haze(ASTM D1003/1044,ISO13468),Transmittance T(ISO),Transmittance T(ASTM)
Observer Angle
Illuminant D65,A,C
Displayed Data PASS/FAIL Result
Measuring Time About1.5s
Haze resolution 0.01unit
Haze repeatability Φ20mm,within 0.1(After preheating and calibration of the instrument, standard deviation values of standard haze tablets with a test haze of about 40 for 30 times were taken at intervals of 5s.)
Inter-instrument Error Φ20mm,with 0.4(After the instrument is preheated and calibrated, the standard deviation between the standard haze tablet and the reference value is tested at an interval of 5s.)
Dimension L*W*H=290X211X511mm
Weight About7.6kg
Power AC 24V, 3A Power adapter power supply
Illuminant Life Span 5 years, more than 3 million times measurements
Display 7-inch TFT color LCD, Capacitive Touch Screen
Data Port USB & Print serial port
Data Storage Standard 1000 pcs,sample 20000 pcs
Language Simplified Chinese,Traditional Chinese,English
Operating Environment 0~40℃(32~104°F)
Storage Environment  -20~50℃(-4~122°F)
Standard Accessory Power Adapter, User Guide,Quality Management Software (download from the official website),USB cable,0% calibration box,Aperture.
Optional Accessory Micro-printer,Test fixture, standard haze sheet, foot switch
Note: The specifications are subject to change without notice.