Features :

1. Fast and accurate—the required measurement time is extremely short, usually tens of seconds to give the content result. With high reproducibility and reproducibility.
2. Low detection limit—adopts EDXRF technology, low background and thin beryllium window proportional counter tube, secondary filter. Reach the lower detection limit.
3. Small and light—only about 5kg (including battery, printer), atmospheric atmosphere measurement without auxiliary equipment, especially suitable for on-site rapid detection.
4. Long-term continuous use—it has extremely low power consumption, using lithium-ion batteries, can be used continuously for a long time.
5. Strong adaptability—automatic C/H ratio correction, matrix effect correction, suitable for a variety of samples. Automatic working curve switching, suitable for wide measuring range.
6. Long-term stability—adopts variable gain digital multi-channel, PHA automatic adjustment, deviation correction and other functions, with excellent long-term stability.
7. Environmental protection and energy saving—radiation protection meets exemption requirements. No contact or damage to the sample during analysis, no pollution, no chemical reagents, and no burning.
8. Easy operation—touch screen operation. The sample is directly loaded into the sample cup, and after putting it into the instrument, you only need to press the [Start] key to truly realize one-key operation.
9. High reliability—integrated design, high degree of integration, strong environmental adaptability, strong anti-interference ability, and high reliability.
10. Data transfer and storage—it can store massive data, support data retention after power failure, and can be viewed at any time. Also it can transfer the stored data to the PC.
11. High cost performance-No need for gas, vacuum, dilution, and extremely low operation and maintenance costs. The price is rather competitive.

Specifications :

  • GD-P4294
  • GOLD
  • GD-P4294
X-ray tube Voltage: ≤10keV,Current: ≤1mA, Power: ≤10W
Detector Ultra-thin beryllium window proportional counter tube
Detection limit (300s) 2.6ppm
Measuring range 3 times the detection limit~9.99%
Linearity error /Analysis accuracy Meet the relevant requirements of GB/T 17040, GB 17411, ASTMD4294, etc.
System analysis time 1~999s, recommended value: 300s or 600s for micro measurement, 100s for constant measurement
Operational environment Temperature -10℃~40℃, relative humidity ≤95% (35℃)
Power supply 20.8Ah, 11.1V built-in lithium-ion battery pack, can work for 5 hours under measurement
Power consumption ≤50W
Dimensions and weight 242mm (W)×258mm (D)×142mm (H),5.6kg(including battery and printer)